An air shower is a small booth at a cleanroom entrance, fitted with nozzles in the walls and sometimes the ceiling. When a person steps in and the doors close, fans push HEPA-filtered air through the nozzles at high speed for a short timed cycle while the person turns slowly. Air returns through low-level grilles, passes through filters and recirculates. The doors are interlocked so they cannot both be open, which also lets the booth act as an airlock.
An air shower can remove loose particles sitting on garment surfaces. It cannot make up for poor gowning, dirty garments, or skin and hair shedding under a garment. Its value depends on the industry:
- Common in electronics, semiconductor and some manufacturing cleanrooms with heavy particle loads.
- Less common in pharmaceutical and pharmacy cleanrooms, where gowning procedure and pressure cascades do most of the work.
Planning points:
- Throughput: a cycle per person adds up at shift change. Size for peak traffic or add a second unit.
- Access: carts and wheelchairs may need a separate route.
- Egress: confirm emergency door release and an alternative exit with the building and fire authority.
- Noise: fans and nozzles are loud for the occupant during the cycle.
See air showers: when they help and the air shower.