Fume hood testing and certification

Fume hood testing to ANSI/ASHRAE 110, face velocity checks, airflow monitor calibration and annual certification reports for Canadian labs.

A fume hood protects people only if it contains what is released inside it. Testing shows whether it does, under the conditions in your lab, and finds problems before they cause an exposure. Our fume hood testing service covers new hoods at commissioning, existing hoods on an annual cycle, and hoods that have been moved or modified.

What the service covers

  • Containment testing. Tests to ANSI/ASHRAE 110, which combines face velocity measurement, smoke visualization and tracer gas release with a mannequin standing in front of the sash. Tests can be done as manufactured, as installed or as used. Our ASHRAE 110 guide explains each part.
  • Face velocity checks. A grid of readings across the sash opening at the set working height. Face velocity is useful, but it does not prove containment on its own; see what face velocity measures.
  • Routine certification. Periodic checks in line with CSA Z316.5, ANSI/AIHA Z9.5 and your organization's program. Annual testing is the common interval.
  • Airflow monitors. Calibration and alarm checks for hood monitors, so users can trust the display.
  • Variable air volume response. Checks of how quickly VAV hoods react when the sash moves.
  • Room conditions. Cross-draft measurements near the hood face, where supply diffusers, doors and traffic can disturb containment.
  • Hood labels. Labels showing the test date, result and the sash height used.
  • Biosafety cabinets. Biosafety cabinets need field certification to NSF/ANSI 49 or its CSA or EN equivalent by a qualified certifier. We can schedule that alongside hood testing.

Acceptance criteria come from the owner, its EHS team and the engineer of record. Where results point to changes in building systems, any stamped design is prepared by a professional engineer licensed in your province.

How it works

  1. Inventory and scope. We list every hood with its type, size, location, controls and test history, and agree which tests each one needs.
  2. Pre-test review. We check design flows, sash positions and acceptance criteria with your EHS team or engineer.
  3. Visual inspection. We look at the sash, baffles, airfoil sill, liner, lights, monitor and services.
  4. Face velocity. We measure the grid at the working sash height and compare it with the design value.
  5. Smoke visualization. We watch airflow at the face, sill and work surface for reverse flow or escape.
  6. Tracer gas. Where in scope, we release tracer gas and measure leakage at the breathing zone of a mannequin.
  7. Monitor calibration. We set and check the airflow monitor and its alarm.
  8. Report and retest. We issue the report, label each hood and retest after corrections.

What you receive

  • ASHRAE 110 test reports for hoods tested to that method
  • Face velocity data for each hood
  • Hood inventory with locations, sizes, types and test dates
  • Deficiency list with likely causes and suggested corrections
  • Airflow monitor calibration records
  • Calibration records for our test instruments
  • Labels fixed to each hood

Typical projects

  • Commissioning new hoods before a lab opens, as covered in commissioning and annual certification
  • Annual testing of every hood on a campus or across a company's sites
  • Testing after a hood is relocated or its exhaust system changes
  • Investigating complaints of smells or alarms at a specific hood
  • Checking hoods before and after a VAV conversion

Good to know

A hood that passes in an empty lab can fail when a supply diffuser blows across its face or when users raise the sash above the tested height. Testing as installed, and sometimes as used, gives the most realistic picture.

To schedule testing or plan a certification program, request a consultation.