A HEPA filter is a deep, pleated sheet of fine glass-fibre media sealed into a rigid frame. It traps particles through a combination of interception, impaction and diffusion. Particles around 0.3 µm are close to the hardest size to catch, so a filter rated at that size performs better on both smaller and larger particles.
Two rating systems are in common use:
- North American practice defines HEPA as at least 99.97% efficient at 0.3 µm.
- EN 1822 and ISO 29463 rate filters at their most penetrating particle size. Class H13 is at least 99.95% efficient and H14 at least 99.995%.
HEPA filters are the basis of biosafety cabinets, clean benches, fan filter units and cleanroom supply systems. They do not remove gases or vapours, which is one reason a biosafety cabinet is not a chemical fume hood.
A filter is only as good as its seal. After installation and periodically afterwards, certifiers scan the filter face and frame with an aerosol challenge and a photometer or particle counter to find leaks. Damaged media and gasket leaks are common findings. Handle filters by the frame, never touch the media, and keep protective grilles in place.
For higher efficiency, see ULPA filter. For testing, see cleanroom certification and monitoring.