An interlock links two or more doors so that while one is open, the others stay locked. It enforces the sequence an airlock depends on: in, close, then out.
Door interlocks usually combine electromagnetic locks or electric strikes, door position switches and a controller. Indicator lights at each door show when it is safe to open. Pass-through chambers often use simpler mechanical interlocks, and air showers have their own built-in interlock controls.
Typical uses:
- Personnel and material airlocks between cleanroom classes or grades.
- Gowning room sequences.
- Pass-through chambers and air showers.
- Entries to containment labs.
Interlocks can include a time delay before the second door releases, giving the pressure in the airlock time to recover.
Life safety comes first. An interlock must never trap people. Systems need an emergency release, such as an override button or break-glass device, and a defined response to power failure and fire alarm, usually releasing the locks. These requirements come from the building code and the local fire authority, so coordinate interlock design with the fire alarm and access control systems from the start.
See cleanroom doors and interlocks and pressure cascades and airlocks.