The airfoil sill, often just called the airfoil, is the curved or angled member mounted across the front of a fume hood at the leading edge of the work surface. Air entering the hood flows smoothly over its rounded profile instead of tumbling over a square edge. A gap underneath it lets a sheet of air sweep across the work surface toward the rear baffle.
Without an airfoil, air separates at the front edge and forms a rolling eddy just inside the sash plane. Vapours released near the front can be caught in that eddy and carried out toward the user. The airfoil reduces this, which is one reason modern hoods contain better at moderate face velocities than older square-edged designs. In many designs the gap stays open when the sash is fully lowered, so a closed hood still draws some air across the work surface.
Good practice:
- Keep the airfoil clear. Do not rest bottles, tubing, cords or paper on it or in the slot below it.
- Set work at least 150 mm (6 in) behind the sash plane.
- Route cords only through openings the manufacturer provides for them.
- Check that the airfoil is back in place after cleaning or maintenance.
Learn more in why airfoil sills matter, or see the benchtop bypass hood for a typical arrangement.