For a fume hood, containment means capturing what is released inside the hood and keeping it out of the user's breathing zone. It is the outcome that matters. Face velocity, airfoils, baffles and sash design are all means to that end.
Containment can be measured directly. The ANSI/ASHRAE 110 tracer gas test releases a gas inside the hood and measures how much reaches a mannequin's breathing zone. A hood with a good face velocity can still leak when conditions are poor. Common causes:
- Cross drafts from supply diffusers, open doors or people walking past.
- A sash raised above its working height.
- Large equipment near the front, or blocking the baffle slots.
- Hot processes that overwhelm the hood's airflow.
- Quick arm movements and working too close to the sash plane.
The word has broader uses. In biosafety, primary containment refers to equipment such as biosafety cabinets plus safe practice, and secondary containment refers to the room and building design. In Canada, the containment level of a facility (CL1 to CL4) is set under the Canadian Biosafety Standard. Cleanrooms for hazardous drugs use negative pressure to provide containment as well as cleanliness.
See how a laboratory fume hood works and cross drafts, diffusers and supply air.