Baffles are panels mounted a short distance in front of the back wall of a fume hood. The space behind them forms a plenum that connects to the exhaust collar. Slots between and around the panels, typically at the top, middle and bottom of the chamber, set where air leaves the work area.
Without baffles, most air would rush straight to the exhaust collar at the top, leaving slow, swirling zones near the work surface where many vapours collect. Properly set slots pull air evenly across the full height and width of the opening. The bottom slot is especially important because it draws heavier-than-air vapours along the work surface.
Some hoods have fixed baffles set at the factory; others let users adjust slot widths, for example opening the top slot for heat-producing work. Many manufacturers recommend leaving adjustable baffles at the factory setting unless the certifier advises otherwise.
Common problems:
- Bottles or equipment pushed against the back, blocking the bottom slot.
- Baffles removed for cleaning and not reinstalled correctly.
- Large apparatus sitting flat on the work surface. Raising it about 50 mm (2 in) on blocks lets air pass underneath.
See baffles and slot adjustment and the fume hood maintenance checklist. Baffles work together with the airfoil sill to give an even face velocity.