Face velocity is the average speed of the air moving into a fume hood through the open sash. A technician measures it with a calibrated anemometer at a grid of points across the opening, usually with the sash at its working height, and averages the readings. Results are given in metres per second or feet per minute.
Design targets for conventional hoods typically fall around 0.4 to 0.5 m/s (80 to 100 fpm). Some high-performance low-flow hoods are designed to operate lower, often around 0.3 m/s (60 fpm), but only where containment testing supports it. More is not better: very high velocities create turbulence at the sash edges and in the wake of the user's body, which can pull vapours out of the hood.
Face velocity is an indicator, not a measure of containment. Cross drafts from doors and diffusers, people walking past, apparatus placed near the sash and the user's own habits all affect what escapes. That is why commissioning normally adds smoke visualization and ANSI/ASHRAE 110 tracer gas tests. The owner's standards and CSA Z316.5 set the acceptance values for a given site.