ANSI/ASHRAE 110

The standard test method for laboratory fume hood performance, combining smoke visualization, face velocity and tracer gas containment tests.

ANSI/ASHRAE 110, Method of Testing Performance of Laboratory Fume Hoods, defines a repeatable way to check how well a hood contains what is released inside it. It is a test method, not a pass/fail standard. Acceptance criteria come from other documents, such as ANSI/AIHA Z9.5, CSA Z316.5 or the owner's own specification.

The test has three main parts:

  • Flow visualization: smoke released at the face and inside the hood to reveal reverse flow, dead zones and escape.
  • Face velocity: averaged readings across the sash opening.
  • Tracer gas containment: sulfur hexafluoride released at 4 L/min from an ejector inside the hood, while a mannequin with a detector at its breathing zone stands at the face.

A sash movement test is often added for VAV hoods to check how well containment holds while the sash moves and airflow adjusts.

Results are written as a rating such as AM 0.05, meaning the measured breathing-zone concentration in ppm at the 4 L/min release. Tests can be run as manufactured (AM, in the factory), as installed (AI, in the lab) or as used (AU, with real equipment in place). Commonly cited acceptance values are 0.05 ppm for AM and 0.1 ppm for AI, but the governing document for your site decides.

When buying a hood, ask for the as-manufactured test report and plan for an as-installed test through a certification service. See ASHRAE 110 containment testing explained.