Tracer gas testing is the core of the ANSI/ASHRAE 110 fume hood test. Instead of inferring performance from airflow readings, it measures containment directly.
The usual tracer is sulfur hexafluoride (SF6). It is chemically inert, low in toxicity and detectable at very low concentrations. In the standard method:
- An ejector inside the hood releases the gas at 4 L/min, at left, centre and right positions 150 mm (6 in) behind the sash plane.
- A mannequin stands at the face of the hood with a detector probe at its breathing zone.
- The detector logs the concentration over a set period, in parts per million.
- The results set the hood's rating, written for example as AM 0.05 for an as-manufactured test.
A rating is only meaningful with its release rate, sash position and test conditions stated. Results also depend on room conditions, which is why as-installed tests are run in the finished lab with the HVAC system operating normally.
SF6 is also a very potent greenhouse gas. Ask your certifier which tracer and detector they use and how they handle it.