LaboratoryEquipment.caCasework · Fume hoods · Cleanrooms Placeholder data
Glossary

Sash stop

A mechanical stop that holds a vertical fume hood sash at its tested working height, with a manual override for loading equipment.

A sash stop is a catch, bumper or spring latch in the sash track that halts a vertical sash at a set working height. The user can release it to raise the sash fully when loading large apparatus, and it re-engages when the sash comes back down.

The stop height is not arbitrary. The exhaust system is sized so the hood reaches its design face velocity with the sash at that opening, and certification tests are run there. A working opening of around 450 mm (18 in) is common, though the right height depends on the hood design and how it was tested. Above the stop, a constant-volume hood's face velocity drops and containment may no longer match the tested result.

Practical points:

  • Mark the certified sash height on the hood with a label or arrow, and check it at each annual certification.
  • Treat the full-open position as a setup position only.
  • On VAV hoods the system is often sized for the stop height, so raising the sash past it can exceed the valve's maximum flow and trigger a low-flow alarm.
  • Do not remove or disable stops because they are inconvenient. Report the problem instead.

See sash height and safe working habits for day-to-day practice, and how the sash itself is built.