ISO 14644-1 classifies cleanrooms and clean zones by the number of airborne particles allowed per cubic metre of air at specified particle sizes. Each class step is a factor of ten. The classes used most often in labs, pharmacy and manufacturing are:
| ISO class | Max particles ≥0.5 µm per m³ | Nearest former FS 209E class |
|---|---|---|
| ISO 5 | 3,520 | Class 100 |
| ISO 6 | 35,200 | Class 1,000 |
| ISO 7 | 352,000 | Class 10,000 |
| ISO 8 | 3,520,000 | Class 100,000 |
The US Federal Standard 209E was cancelled in 2001, but its class names still appear in conversation and older documents.
A class means little without its occupancy state. ISO 14644-1 recognizes three: as-built (empty room), at-rest (equipment installed, no people) and operational (normal work under way). A room that meets ISO 7 at rest may not do so in operation.
Classification is confirmed by a particle count at a number of locations set by the room's area. Pharmaceutical work also uses the GMP grades A to D, which relate to ISO classes but add microbial limits.
See ISO 14644-1 cleanroom classes explained and cleanroom design-build.