Laminar flow is the long-established name for clean air that moves in parallel streams at a steady velocity from a HEPA filter across a work zone. ISO 14644 now uses the term unidirectional airflow instead, because at room scale the flow is not strictly laminar in the fluid-mechanics sense. Both terms appear in catalogues and specifications.
The most familiar laminar flow device is the clean bench, often called a laminar flow hood:
- Horizontal flow: filtered air blows from the back of the bench toward the user.
- Vertical flow: filtered air blows down from the top onto the work surface.
Air velocity at the filter face is commonly around 0.45 m/s (90 fpm).
A clean bench protects the product, not the person. A horizontal bench blows air, and anything released into it, straight at the user's face. Never use a clean bench for infectious materials, hazardous drugs, toxic powders or volatile chemicals. Use a Class II biosafety cabinet or a fume hood for those.
Good practice in any laminar flow zone: keep items out of the path between the filter and the critical work, avoid rapid movements, and do not block the filter face or return paths.
See clean benches and PCR workstations and the horizontal laminar flow clean bench.