A Class II biosafety cabinet provides three kinds of protection at once:
- Personnel: room air flows in through the front opening, away from the user.
- Product: HEPA-filtered air flows down over the work surface, protecting cultures and samples from contamination.
- Environment: exhaust air passes through a HEPA filter before it leaves the cabinet.
Type A2 is defined in NSF/ANSI 49. Its key features include a minimum average inflow velocity of 0.51 m/s (100 fpm) and recirculation of roughly 70% of the cabinet's air, with about 30% exhausted. Contaminated plenums are under negative pressure or surrounded by negative-pressure plenums. The exhaust can discharge into the room or connect to the building exhaust through a canopy (thimble) connection, which makes the cabinet less sensitive to fluctuations in the building system.
An A2 cabinet is not a chemical fume hood. HEPA filters do not capture vapours, and recirculated air would carry them back into the work zone. Work with volatile chemicals needs a different cabinet type or a fume hood; any small exceptions for canopy-connected cabinets are for the biosafety officer to approve.
Cabinets are field certified after installation, after relocation and at least annually, commonly to NSF/ANSI 49 or the equivalent CSA standard.