Class II, Type A2

The most widely used biosafety cabinet type, protecting the user, the work and the room with inflow air and HEPA-filtered airflow.

A Class II biosafety cabinet provides three kinds of protection at once:

  • Personnel: room air flows in through the front opening, away from the user.
  • Product: HEPA-filtered air flows down over the work surface, protecting cultures and samples from contamination.
  • Environment: exhaust air passes through a HEPA filter before it leaves the cabinet.

Type A2 is defined in NSF/ANSI 49. Its key features include a minimum average inflow velocity of 0.51 m/s (100 fpm) and recirculation of roughly 70% of the cabinet's air, with about 30% exhausted. Contaminated plenums are under negative pressure or surrounded by negative-pressure plenums. The exhaust can discharge into the room or connect to the building exhaust through a canopy (thimble) connection, which makes the cabinet less sensitive to fluctuations in the building system.

An A2 cabinet is not a chemical fume hood. HEPA filters do not capture vapours, and recirculated air would carry them back into the work zone. Work with volatile chemicals needs a different cabinet type or a fume hood; any small exceptions for canopy-connected cabinets are for the biosafety officer to approve.

Cabinets are field certified after installation, after relocation and at least annually, commonly to NSF/ANSI 49 or the equivalent CSA standard.