A strong acid that becomes a powerful oxidizer when hot, and whose vapours can leave explosive perchlorate deposits in ordinary hoods and ducts.

Perchloric acid (HClO4) is used for wet digestions in metals analysis, mining assay, environmental and materials labs. At room temperature, the dilute commercial acid behaves mainly as a strong, corrosive acid. Heated, it becomes a powerful oxidizer, and its vapours can condense in hood interiors, ductwork and fans. There they can react with metals and organic material to form perchlorate deposits that are sensitive to shock, friction and heat.

For that reason, work that heats perchloric acid is commonly restricted to a dedicated perchloric acid hood with:

  • A seamless, coved liner and work surface in a resistant material, commonly stainless steel.
  • An integral washdown system to rinse the baffle, plenum, duct and fan.
  • Materials in the airstream chosen to avoid reactive organics such as ordinary plastics, gaskets and sealants.
  • Its own dedicated, straight, vertical duct and fan, never shared with other hoods.
  • Clear labelling so the hood is not used for organic chemicals.

Old hoods and ducts with a history of perchloric acid use need specialist inspection and testing before anyone opens, modifies or removes them.

Whether a specific procedure needs a perchloric hood is a decision for the organization's EHS office. See perchloric acid hoods and washdown systems, the perchloric acid hood and the mining and geochemistry assay lab.