A washdown system sprays water through nozzles placed behind the baffle, in the exhaust duct and often at the fan. The water flushes condensed residues off these surfaces and drains back down to a trough at the rear of the hood's work surface, then to a drain.
Washdown is a standard feature of perchloric acid hoods, where it prevents shock-sensitive perchlorate deposits from building up. It is also used on some hoods for other heated acid digestions that leave corrosive or crystalline deposits.
Users typically run the washdown after each session of perchloric acid work, following the lab's written procedure. Systems may use a manual valve or a timer.
Design points that make washdown work:
- Vertical, smooth, corrosion-resistant ductwork with no horizontal runs where water and residue can pool.
- A drain sized for the spray flow, with chemical-resistant piping.
- A water supply with backflow protection, such as a vacuum breaker or a backflow preventer, as the plumbing design requires.
- Easy access to nozzles for inspection, since blocked nozzles leave unwashed areas.
Washdown water may carry acids and metals. The organization's EHS office and local sewer bylaws decide whether it can go to drain, and what treatment it needs first.
See perchloric acid hoods and washdown systems and the perchloric acid hood.