A ductless fume hood or filtered enclosure uses its own fan to pull air in through the front opening, then through one or more filters, and discharges the cleaned air back into the lab. The filters are matched to the hazard:
- Activated carbon adsorbs many organic vapours. Impregnated carbons target specific groups such as acid gases, ammonia or formaldehyde.
- HEPA filters capture particles such as powders, but not vapours.
The appeal is simple: no ductwork, no roof fan, no makeup air to condition, and units can often be moved. That makes them useful for small, well-defined tasks such as weighing powders, staining, or handling known solvents in modest amounts.
The limits are just as real. A carbon filter has finite capacity, holds some chemicals poorly, and can saturate and release vapour without visible warning. Ductless units are generally unsuitable for unknown or changing chemical use, large volumes, highly toxic materials, heated acid digestions and perchloric acid.
Good practice:
- Get the manufacturer's written confirmation that each chemical and quantity is suitable for the filter.
- Use saturation detection and a filter-change schedule, and keep a log.
- Have EHS approve the unit for its specific use.
See ductless filtered fume hoods: uses and limits and the ductless filtered fume hood.