A kilo lab sits between the research bench and the plant. Process chemists and engineers scale reactions from grams to kilograms, make material for toxicology studies or early clinical supply, and prove out a process before it moves to a pilot plant or a contract manufacturer. Pharmaceutical and fine chemical companies, contract manufacturers and some universities run them. The equipment is laboratory glass at large scale: jacketed reactors of roughly 10 to 100 L (about 3 to 26 US gal), large rotary evaporators, filters and dryers, all inside walk-in hoods.
Hazards and containment
Scale changes the risk. A runaway reaction, a cracked reactor or a solvent spill involves litres, not millilitres. The main hazards are:
- Large volumes of flammable solvents, and the static charge that builds up when they are pumped or poured
- Exothermic reactions and pressure build-up
- Toxic or potent compounds during charging, sampling and filtration
- Heavy lifting of drums and vessels
- Very cold and very hot heat-transfer fluids from chillers and thermostats
Every operation with open solvent belongs inside a walk-in fume hood, with the sashes closed while reactions run. Ground and bond reactors, drums and transfer lines. Charge potent solids in a ventilated enclosure or through a closed transfer system. Each process should go through a hazard review before it runs at scale.
Equipment and casework
- Walk-in hoods, commonly 1.8 to 2.4 m (6 to 8 ft) wide, with reactor frames standing on the floor inside
- Chillers and thermostats beside or behind the hoods
- A benchtop hood for weighing, sampling and small reactions
- Instrument benches for in-process testing by HPLC and GC
- Flammable cabinets for working stock, with bulk solvent kept in a rated storage room
- A corrosives cabinet on a different wall
- A work-up island and mobile tables for staging carts and drums
Layout principles
Give the walk-in hoods the long wall farthest from the door, and leave a wide clear zone in front of them: at least 1.5 m (5 ft), and more where reactors and drums are wheeled in. Walk-in hoods open down to the floor, so supply air and foot traffic near them affect containment more than they would at a benchtop hood. Keep the route from the door to the hoods clear for equipment moves, and use double doors wide enough for the largest reactor frame.
Place in-process instruments and paperwork stations away from the hoods. Keep flammable storage close to the hoods but out of the exit path, and keep corrosives on another wall. Put the emergency shower and eyewash near the exit and within 10 seconds of every hood. Ceiling height matters too: walk-in hoods and tall reactor frames need a ceiling of at least 3 m (10 ft), plus space above for ducts.
Finishes and worktops
Epoxy resin worktops resist solvents, acids and heat at benches and sinks. Walk-in hoods have no worktop, so the choice there is the liner: fiberglass for solvent and general work, or stainless where the process calls for it. Painted steel casework is durable and adds little fuel to a fire. Floors should be chemical-resistant and seamless, and static dissipative where solvents are transferred.
Utilities and services
Kilo labs are service-heavy:
- High exhaust volumes for several walk-in hoods, which drive large make-up air and energy costs; see fume hood energy use
- Nitrogen for inerting, plus vacuum and compressed air at each hood
- Chilled and cooling water, or space and power for local chillers
- Three-phase power, with electrical equipment rated for the area classification your engineer assigns
- Grounding points at each hood and drum station
- Drains with spill containment so a spill cannot reach the sewer
Compliance points in Canada
The National Fire Code of Canada, NFPA 30 and NFPA 45 limit flammable liquid quantities and set storage and fire protection rules. A kilo lab often reaches those limits, which is why bulk solvent lives elsewhere. Walk-in hoods should be certified against CSA Z316.5 and tested to ANSI/ASHRAE 110. If the lab makes material for clinical use, good manufacturing practice expectations shape cleaning, documentation and finishes. The engineer of record and the local fire authority have the final say on classification and quantities.
About the starting layout
The 3D starting layout is a 14 by 10 m (46 by 33 ft) room with a 3.6 m (11 ft 10 in) ceiling, three 2.4 m walk-in hoods along the back wall and a fourth on the side, a benchtop hood for sampling, in-process instrument benches, two flammable cabinets, a corrosives cabinet on another wall, a wide double doorway and a work-up island with a staging table. Adapt it in the lab planner or book a consultation to size hoods and exhaust for your process.
Starting layout: bill of materials
30 items in a 14 × 10 m room with a 3.6 m ceiling. Budgetary list prices; adjust anything in the planner.
| Code | Product | Qty | Unit |
|---|---|---|---|
| AB-CF-1800 | Adaptable C-frame bench Painted steel, Arctic white · Epoxy resin, black | 2 | $5,360 |
| BC-4D-0900 | Base cabinet, 4 drawers Painted steel, Arctic white · Epoxy resin, black | 3 | $2,270 |
| BC-DD-0900 | Base cabinet, drawer over doors Painted steel, Arctic white · Epoxy resin, black | 2 | $1,900 |
| BC-HD-0900 | Base cabinet, hinged doors Painted steel, Arctic white · Epoxy resin, black | 3 | $1,760 |
| BC-KS-1200 | Knee space with modesty panel Painted steel, Arctic white · Epoxy resin, black | 1 | $1,110 |
| BC-SK-1200 | Sink base with epoxy sink & faucet Painted steel, Arctic white · Epoxy resin, black | 1 | $3,180 |
| ES-CB-0600 | Emergency shower & eyewash | 1 | $1,850 |
| FH-BP-1800 | Benchtop fume hood, bypass Painted steel, Arctic white · Fiberglass liner | 1 | $12,100 |
| FH-WI-1800 | Walk-in fume hood Fiberglass liner | 1 | $21,500 |
| FH-WI-2400 | Walk-in fume hood Fiberglass liner | 3 | $24,800 |
| IB-RS-3600 | Island bench with reagent shelving Painted steel, Arctic white · Epoxy resin, black | 1 | $17,280 |
| MT-C-1800 | Mobile lab table Epoxy resin, black | 1 | $2,180 |
| SC-CR30-1090 | Corrosives cabinet, 30 gal | 1 | $2,290 |
| SC-FL45-1090 | Flammable liquids cabinet, 45 gal | 2 | $2,480 |
| TC-ST-0900 | Tall storage cabinet Painted steel, Arctic white | 3 | $2,050 |
| WC-GL-0900 | Wall cabinet, glass doors Painted steel, Arctic white | 1 | $1,080 |
| WC-GL-1200 | Wall cabinet, glass doors Painted steel, Arctic white | 1 | $1,290 |
| WC-SD-0900 | Wall cabinet, solid doors Painted steel, Arctic white | 1 | $950 |
| WC-SD-1200 | Wall cabinet, solid doors Painted steel, Arctic white | 1 | $1,140 |
| Starting layout total | $178,070 |